PERFORMANCES | |
Calibrated ranges1 |
Density: 600 kg.m-3 … 1500 kg.m-3 Viscosity: 0.3 cP … 400 cP (500 cSt) Water activity: 0 aw … 1 aw Temperature: 0 °C … 85 °C |
Accuracy2 (typical, max) |
Density: 0.1 %F.S. (typ) 0.5 %F.S. (max) Viscosity: 0.8 %F.S. (typ) 2 %F.S. (max) Water activity: 1 %F.S. (typ) 3 %F.S. (max) Temperature: 0.2 °C 0.5 °C |
Repeatability2 |
Density: 0.1 %F.S. Viscosity: 0.5 %F.S. Water activity: 0.5 %F.S. Temperature: 0.2 °C |
SPECIFICATIONS | |
Real-time calibrated output | Density (kg.m-3), Viscosity (cP – cSt), Water activity (aw), Temperature (°C) |
Operating conditions |
Temperature: -40 °C … 125 °C Pressure: Atm … 10 bar (burst pressure 50 bar) |
Communication | 4 – 20mA, USB, RS232, ModBus TCP |
Acquisition frequency (max) | 1 Hz |
Response time | < 2 sec |
Sealing ring | Viton ©, Kalrez © (option) |
Wetted parts | Stainless steel 316L, Inconel, Hastelloy (option) |
Weight | 320 g |
Dimensions |
38mm x 38mm x 40mm (NeSSI design) Dimensions can be modified for OEM applications Several implementation proposals are shown in the “documentation” section |
Certifications | CE, ATEX – II 1G Ex ia IIC T4 |
1. The density era is divided into meaningful ranges: 600 – 950 // 900 – 1200 // 1100 – 1500
The viscosity era is divided into meaningful ranges: 0.3 – 10 // 3 – 80 // 10 – 200
2. Indicated performances have been obtained in controlled laboratory conditions using sensor’s digital output. In practice, additional inaccuracy might arise due to application’s conditions, use of the analog mode, fluid characteristics…