PERFORMANCES | |
Calibrated ranges1 |
Density: 30 g.m-3 … 8000 g.m-3 Molar mass: 2 g.mol-1 … 50 g.mol-1 Temperature: -25°C … 90°C |
Accuracy2 (typical, max) |
Density: typ: 2,5 g.m-3 + 0.08 %reading max: 5 g.m-3 + 0.2 %reading Molar mass: 0.8 %F.S. (typ) 2 %F.S. (max) Temperature: 0.2 °C 0.5 °C |
Repeatability2 |
Density: 0.01 %F.S. Molar mass: 0.1 %F.S. Temperature: 0.2 °C |
SPECIFICATIONS | |
Real-time calibrated output |
Density (g.m-3), Molar mass (g.mol-1), Temperature (°C) AS an option, Sphere® sensor: Pressure (Pa) |
Operating ranges |
Temperature: -40°C … 95 °C Pressure: Atm … 8 bar (operating up to 40 bar) |
Communication | 4 – 20mA, USB, RS232, ModBus TCP |
Mechanical integration (other on demand) |
By-pass: I/O 1/8″ NPT, NeSSI Inline, to be screwed: G 1/2″ |
Flow velocity (max) |
< 1m.sec-1 |
Acquisition frequency (max) | 1 Hz |
Response time | < 1 sec |
Sealing ring | Viton ©, Kalrez © (option) |
Wetted parts | Stainless steel 316L, Inconel (option), Hastelloy (option) |
Weight | 460 g |
Dimensions |
160mm x 40mm x 40mm (including Sphere® sensor) Dimensions can be modified for OEM applications |
Certifications | CE, ATEX – II 1G Ex ia IIC T4 |
1. The density era can be divided into smaller ranges for improved performances
2. Indicated performances have been obtained in controlled laboratory conditions using sensor’s digital output. In practice, additional inaccuracy might arise due to application’s conditions, use of the analog mode, fluid characteristics…